Our Calibration Services

Vacuum or Atmospheric Furnace Calibration
01

Vacuum or Atmospheric Calibration

  • TUS (Temperature Uniformity Survey)
  • SAT (System Accuracy Test)
  • Resident / Non-Resident / Comparison
  • Compliance with AMS2750 / NADCAP / AQI / CQI-9
Thermal or Atmospheric Equipment Calibration
02

Thermal or Atmospheric Equipment Calibration

  • Furnaces, ovens, heat treaters
  • Thermocouple & temperature calibration
  • Compliance with AMS2750 / NADCAP / AQI / CQI-9
  • Sub-zero chambers up to -100 °C
Controller Calibration
03

Controller Calibration

  • Temperature & process controllers
  • PID tuning
  • Multi-channel recorder accuracy checks
Vacuum Sensor Calibration
04

Vacuum Sensor Calibration

  • Pirani, piezo/Pirani, ionization & capacitance manometer transducers
  • Absolute & relative pressure, rough to high vacuum
  • Full-scale range accuracy check
  • Traceable to national standards
Pressure & Compound Gauge Calibration
05

Pressure & Compound Gauge Calibration

  • Analog & digital pressure gauges
  • Compound gauge accuracy tests
  • Ranges: vacuum to 500 bar
ISO 17025 Traceable AMS2750 Compliance NADCAP Ready CQI-9 Compliance On-Site Service In-Lab Calibration Fast Turnaround Certified Technicians Calibration Certificates

Why Choose Us

Our Advantages

  • ISO 17025 traceability (sub-contract)
  • Fast turnaround times
  • Calibration certificates provided
  • Experienced, certified technicians

Service Highlights

  • Vacuum & atmospheric furnaces
  • Thermal equipment & sub-zero chambers
  • Controllers
  • Vacuum sensors
  • Pressure gauges

Add-on Services

  • Equipment maintenance
  • Emergency calibration
  • Validation support
  • Documentation audits

Why Calibration Decides the Quality of Every Cycle

Your furnace only does what the instruments say it is doing. If the thermocouples, controller or vacuum sensors are reading a little high or a little low, the cycle follows the error — and the evidence shows up later as hardness spread, case depth that misses the mark, longer cycle times, or parts that behave differently in service.

Calibration is how you close that gap. It establishes that what the controller reports is what the furnace is actually doing, on the equipment you run, at the temperatures and pressures you use — not on a datasheet. For heat treatment plants working to customer specifications or aerospace and automotive standards such as AMS2750, NADCAP and CQI-9, it is also the record that proves the process was under control.

Lakshmi Vacuum’s calibration division works on the furnaces and instruments we build and service every day, covering vacuum and atmospheric furnaces, thermal equipment, sub-zero chambers, controllers, vacuum sensors and pressure gauges — on-site at your plant or in our lab, with ISO 17025 traceability on a sub-contract basis and certificates you can file.

Calibration Scope at a Glance

ScopeWhat is checkedWhere it is done
Vacuum or Atmospheric CalibrationTUS (Temperature Uniformity Survey) · SAT (System Accuracy Test) · resident / non-resident / comparison · AMS2750 / NADCAP / AQI / CQI-9 complianceOn-site only
Thermal or Atmospheric Equipment CalibrationFurnaces, ovens and heat treaters · thermocouple and temperature calibration · AMS2750 / NADCAP / AQI / CQI-9 compliance · sub-zero chambers up to -100 °COn-site or in-lab
Controller CalibrationTemperature and process controllers · PID tuning · multi-channel recorder accuracy checksOn-site or in-lab
Vacuum Sensor CalibrationPirani, ionization and capacitance manometers · full-scale range accuracy checkIn-lab or on-site
Pressure & Compound Gauge CalibrationAnalog and digital pressure gauges · compound gauge accuracy tests · vacuum to 500 barIn-lab or on-site

Standard Colour Codes for Thermocouples, Wire and Connectors

Colour Coding Extension Grade Alloy Combination Comments for Environment (Bare Wire) Maximum T/C Grade Temp. Range Plug & Jack
(+) Lead(−) Lead
J +− Iron(Magnetic) Constantan(Copper-Nickel) Reducing, vacuum, inert. Limited use in oxidizing at high temperatures. Not recommended for low temperatures. 0 to 750 °C Black
K +− Nickel-Chromium Nickel-Aluminium(Magnetic) Clean oxidizing and inert. Limited use in vacuum or reducing. Wide temperature range, most popular calibration. 0 to 1250 °C Yellow
T +− Copper Constantan(Copper-Nickel) Mild oxidizing, reducing, vacuum or inert. Good where moisture is present. Low temperature and cryogenic applications. -250 to 350 °C Blue
E +− Nickel-Chromium Constantan(Copper-Nickel) Oxidizing or inert. Limited use in vacuum or reducing. Highest EMF change per degree. -200 to 900 °C Purple
N +− Nicrosil(Nickel-Chromium-Silicon) Nisil(Nickel-Silicon-Magnesium) Alternative to Type K, more stable at high temperature. -270 to 1300 °C Orange
R +− PlatinumRhodium-13% Platinum Oxidizing or inert. Do not insert in metal tubes. Beware of contamination. High temperature. 0 to 1600 °C Green
S +− PlatinumRhodium-10% Platinum Oxidizing or inert. Do not insert in metal tubes. Beware of contamination. High temperature. 0 to 1450 °C Green
B +− PlatinumRhodium-30% PlatinumRhodium-6% Oxidizing or inert. Do not insert in metal tubes. Beware of contamination. High temperature. Common use in glass industries. 0 to 1800 °C White(Uncompensated)
C +− TungstenRhenium-5% TungstenRhenium-26% Vacuum, inert, hydrogen. Beware of embrittlement. Not practical below 399 °C. Not for oxidizing atmosphere. 0 to 2300 °C Red

Colour conventions differ between standards (ANSI, IEC, BS), so confirm the standard your plant and customer follow. Thermocouples, compensating cables and connectors: thermocal@lakshmivacuum.com.

When You Need a Calibration

01

New equipment goes in

02

A hot zone, sensor or controller is replaced

03

Your schedule comes round

04

Results start to drift

05

An audit is coming

06

The process changes

How a Calibration Visit Works

1. Tell us what needs calibrating

2. We agree scope and schedule

3. Calibration is carried out

4. You receive the certificate

Calibration Division — FAQ

Which equipment can you calibrate?

Vacuum and atmospheric furnaces (TUS and SAT); thermal equipment such as furnaces, ovens and heat treaters, including sub-zero chambers up to -100 °C; temperature and process controllers; vacuum sensors including Pirani, ionization and capacitance manometers; and pressure and compound gauges from vacuum up to 500 bar.

Can the calibration be done at our plant?

Yes. Furnace calibration (TUS and SAT) is carried out on-site at your plant. Thermocouples, controllers, sensors and gauges can be calibrated on-site or in our lab, depending on the equipment, the accuracy required and your production schedule.

What traceability and standards apply?

The service is offered with ISO 17025 traceability on a sub-contract basis, aligned with AMS2750, NADCAP, AQI and CQI-9 requirements — the standards your heat treatment customers and auditors ask about.

Do we receive a calibration certificate?

Yes. Calibration certificates are provided for the work carried out, giving you the record your quality system needs.

How quickly can you attend?

Fast turnaround is part of the service, and emergency calibration is available as an add-on when a furnace cannot wait. Scheduling is confirmed as soon as the enquiry is reviewed.